RANCANG BANGUN SISTEM NON-INVASIF PENDETEKSI KADAR ASAM URAT BERBASIS ARRAY NEAR-INFRARED (NIR) DAN REGRESI PLS

RAHMAH, GST. ADELIA RANCANG BANGUN SISTEM NON-INVASIF PENDETEKSI KADAR ASAM URAT BERBASIS ARRAY NEAR-INFRARED (NIR) DAN REGRESI PLS. [Skripsi]

[thumbnail of NASKAH SKRIPSI (GST. ADELIA RAHMAH)_RANCANG BANGUN SISTEM NON-INVASIF PENDETEKSI KADAR ASAM URAT.pdf] Text
NASKAH SKRIPSI (GST. ADELIA RAHMAH)_RANCANG BANGUN SISTEM NON-INVASIF PENDETEKSI KADAR ASAM URAT.pdf

Download (712kB)

Abstract

Conventional uric acid examination is commonly performed using invasive
blood sampling methods, which may cause pain, discomfort, and require medical
personnel as well as specialized equipment. This study aims to design and
implement a non-invasive uric acid detection system based on Near Infrared (NIR)
technology using two TCRT5000 sensors, apply the Partial Least Squares (PLS)
method to develop a uric acid prediction model, and evaluate the system
performance by comparing the predicted results with invasive measurements as the
reference standard. The study employed the Research and Development (R&D)
method using the ADDIE model, consisting of the analysis, design, development,
implementation, and evaluation stages. The proposed system integrates two
TCRT5000 sensors positioned on the fingertip and wrist, an ESP32 microcontroller
for data acquisition and processing, an OLED display for real-time visualization,
and Google Spreadsheet for data storage. The calibration results show that both
sensors have an optimal response at a distance of 2.5 mm with an acquisition time
of 20 seconds. Testing on 50 participants showed that the PLS method successfully
built a predictive model for uric acid levels based on optical signals from both
sensors. The average difference between the non-invasive measurement results and
those from the invasive method was 0.02 mg/dL, with a result acceptance rate of
82%., indicating satisfactory predictive performance. Although the results are
influenced by factors such as sensor placement, skin thickness, and optical signal
quality, the developed system shows potential as an alternative non-invasive uric
acid screening and preliminary monitoring device.
Keywords: uric acid, Near-Infrared (NIR), TCRT5000, Partial Least Squares
(PLS), non-invasive, ESP32.